Specifications | Alice® 3

Number of Neurological Channels: 6 Referential 2 Differential, EOG, EMG
PSG Channels Full set of PSG Parameters w/up to 8Analog Auxiliary Signal Inputs. Automatic analysis provided for SpO2, pH, EtCO2, CPAP, bi-level. Can record up to 28 channels of Data
Input Impedance: 1 MΩ
Input Signal Range: 28µV to 2.5mV
Digital Resolution: 8 bits
Sample Rate: 100Hz
Electrode Impedance Testing: 1Hz, ±0.05 uA Square wave betweenpatient electrodes, capable of measuring 0 to 20K Ohms, user initiated
Calibration Signal: 1 Hz, 250uVpp to amplifier inputswhile Calibration ON
Communication Interfaces: Data sent from Calvin Amplifier toComputer Digital Input Device w/high-speed, standard serial interface.
Digital Filters Available During Analysis: 0.6, 1.0, 2.0, or 4.0 High Pass Filter
20, 30, 40, or 45 Hz Low Pass Filter
50 or 60 Hz Notch Filter

Top of Page

Physical Characteristics:
Size: 8.75 L x 5.75 W x 3.0 H (in)
Weight: 4.02 lb

Top of Page

Power Requirements:
Calvin: 100-250VAC -10%,+6%, 47-63 Hz,
0.65A max at 90VAC

Top of Page

 


Respironics Home  |  Glossary  |   Careers  |  Online Stores  |  Warranties  |  Providers
Suppliers  |  Clinicians  |  Media Resource Library  |  Contact Us
All Rights Reserved  |  Privacy Notice  |  Terms & Conditions  |  Site Map

 